The design of radiology viewing box using charger system and potentiometer

Authors

  • Anak Agung Aris Diartama Postgraduate Imaging Diagnostic Program, Poltekkes Kemenkes Semarang, Semarang Health Polytechnic, Indonesia
  • Susy Suswaty Poltekkes Kemenkes Jakarta II, Jakarta II Health Polytechnic, Indonesia
  • Win Priantoro Poltekkes Kemenkes Jakarta II, Jakarta II Health Polytechnic, Indonesia
  • Sugiyanto Sugiyanto Postgraduate Imaging Diagnostic Program, Poltekkes Kemenkes Semarang, Semarang Health Polytechnic, Indonesia
  • Sudiyono Sudiyono Postgraduate Imaging Diagnostic Program, Poltekkes Kemenkes Semarang, Semarang Health Polytechnic, Indonesia
  • M. Choiroel Anwar Postgraduate Imaging Diagnostic Program, Poltekkes Kemenkes Semarang, Semarang Health Polytechnic, Indonesia
  • Leny Latifah Postgraduate Imaging Diagnostic Program, Poltekkes Kemenkes Semarang, Semarang Health Polytechnic, Indonesia
  • Aris Santjaka Postgraduate Imaging Diagnostic Program, Poltekkes Kemenkes Semarang, Semarang Health Polytechnic, Indonesia
  • Faisal Amri Postgraduate Imaging Diagnostic Program, Poltekkes Kemenkes Semarang, Semarang Health Polytechnic, Indonesia
  • Donny Kristanto Mulyantoro Postgraduate Imaging Diagnostic Program, Poltekkes Kemenkes Semarang, Semarang Health Polytechnic, Indonesia

DOI:

https://doi.org/10.35898/ghmj-1196

Keywords:

Viewing box, Radiology, Potentiometer, Charger system

Abstract

Background: In the process of work to gain the maximum results, a radiologist needs a viewing box tool to read radiographs. Therefore, the authors want to develop a viewing box tool, which in general the work if this tool resembles the factory manufactured tool. The viewing tool box made can adjust the intensity of the light produced.

Objective: to create a viewing box tool by using a potentiometer system.

Methods: This study used applied research method by creating and using the design of viewing box tool by using a potentiometer system and testing the viewing box tool created by using a Lux meter and 15 respondents consisting of five radiologists and 10 radiographers who should fulfill the questionnaire form.

Results: The mean of viewing box illumination reached 220 lux. The results of the questionnaire showed that 100% radiologist gave an A (excellent) and expressed that the viewing box tool created could be used properly and 90% radiographers provided an A (excellent) and expressed that the viewing box tool created could be used properly, while 10% radiographer gave a value of B (moderate).

Conclusion: viewing box tool created could be used properly and obtained optimal results as a tool in reading radiographs. Potentiometer system contained in the viewing box was very helpful in reading radiographs because it allowed to adjust the light intensity according to user needs.

Keywords       :  Viewing box, Potentiometer

Bibliography   : 1980-2011

Downloads

Download data is not yet available.

Author Biography

  • Anak Agung Aris Diartama, Postgraduate Imaging Diagnostic Program, Poltekkes Kemenkes Semarang, Semarang Health Polytechnic, Indonesia

    In this study did not use ethical clearance, due to not using the patient.
    This research is only a tool to help radiologists in reading radiographs
    And in this study no risk to the user

References

REFERENCES

T Nyathi, MSc, AN Mwale, BSc, P Segone, BMedSc, SH Mhlanga, BSc, and ML Pule, BmedSc. Radiographic viewing condition at Johannesburg Hospital. Biomed Imaging Interv J. 2008 Apr-Jun; 4(2):e17

Jenkins, David. Radiographic Photography and Imaging Processes. Roctville: Maryland 20850. 1980

Alter AJ, Kargas GA, Kargas SA, et al. The influence of ambient and viewbox light upon visual detection of low-contrast targets in a radiograph. Invest Radiol. 1982;17(4):402–6.

Chasney,D.N.Radiographic Imaging. Blackwell Scientific Publications: Oxford. 1989

McCarthy E1, Brennan PC. Viewing conditions for diagnostic images in three major Dublin hospitals: a comparison with WHO and CEC recommendations. Br J Radiol. 2003 Feb;76(902):94-7

Kimme-Smith C1, Haus AG, DeBruhl N, Bassett LW. Effects of ambient light and view box luminance on the detection of calcifications in mammography. AJR Am J Roentgenol. 1997 Mar;168(3):775-8.

The Authoritative Dictionary of IEEE Standards Terms (IEEE 100) (edisi ketujuh ed.). Piscataway, New Jersey: IEEE Press. 2000. ISBN 0-7381-2601-2.

Kepmenkes. Pedoman Kendali Mutu (Quality Control) Peralatan Radiodiagnostik. Keputusan Menteri Kesehatan Republik Indonesia Nomor 1250/MENKES/SK/XII/2009: Jakarta. 2009

Downloads

Published

2017-06-22

Issue

Section

Research Articles

Categories

How to Cite

Diartama, A. A. A., Suswaty, S., Priantoro, W., Sugiyanto, S., Sudiyono, S., Anwar, M. C., Latifah, L., Santjaka, A., Amri, F., & Mulyantoro, D. K. (2017). The design of radiology viewing box using charger system and potentiometer. GHMJ (Global Health Management Journal), 1(1), 10-14. https://doi.org/10.35898/ghmj-1196

Plaudit

Share